AP-ML VARIABILITY IN THE PATELLOFEMORAL COMPARTMENT

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 4.



Abstract

Patellofemoral arthroplasty (PFA) has higher revision rates than total knee arthroplasty (TKA) [Van der List, 2015; Dy, 2011]. Some indications for revision include mechanical failure, patellar mal-tracking, implant malalignment, disease progression and persistent pain or stiffness [Dy, 2011; Turktas, 2015]. Implant mal-positioning can lead to decreased patient satisfaction and increased revision rates [Turktas, 2015]. Morphological variability may increase the likelihood of implant mal-positioning. This study quantifies the morphological variability of the anterior-posterior (AP) and medial-lateral (ML) aspects of the patellofemoral compartment using a database of computed tomography (CT) scans.

The analysis presented here used the custom CT based program SOMA (SOMA V.4.3.3, Stryker, Mahwah, NJ). SOMA contains a large database of 3D models created from CT scans. Anatomic analysis and implant fitting tools are also integrated into SOMA to perform morphometric analyses. A coordinate system is established from the femoral head center, the intercondylar notch, and a morphological flexion axis (MFA). The MFA is created by iteratively fitting circles to the posterior condyles and creating and axis through the circles' centers. The sagittal plane is created normal to this axis and through the notch. A coronal plane is created from the femoral head center and the flexion axis. The AP measurement is taken normal to the coronal plane from the anterior cortex sulcus to the intercondylar notch (Figure 1). A 5°-flexed anterior resection is created to run-out at the anterior cortex sulcus. The ML measurement is taken normal to the sagittal plane from the most medial to the most lateral points of the anterior resection (Figure 1). The ML measurements are broken down into medial and lateral components divided by a sagittal plane through the trochlea.

Means and standard deviations of the AP and ML measurements are calculated. The mean and standard deviation for the AP measurement are 24.9mm and 2.8mm, respectively. The data predicts that 99.7% of the population will have an AP measurement between 16.5mm and 33.3mm. The mean and standard deviation for the ML measurement are 54.6 mm and 5.5mm, respectively. The data predicts that 99.7% of the population will have an ML measurement between 38.1mm and 71.1mm A Pearson Correlation value of 0.134 was calculated for AP/ML indicating a very weak positive correlation between the measures. The correlation value and the large measurement ranges indicate that there is high variability between the AP and ML measurements. A scatterplot was created to graphically represent the high variability between the AP and ML width measurements (Figure 2). A Pearson Correlation value of −0.649 was calculated for the medial and lateral components of ML (Figure 3).

The results of this study suggest that patellofemoral morphology is highly variable with respect to the AP and ML dimensions. This variability may impact implant fit and positioning and should be taken into consideration in the design and use of prostheses for PFA.