IN-VIVO BLOOD METAL ION LEVELS IN PATIENTS FOLLOWING TOTAL SHOULDER ARTHROPLASTY

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 3.



Abstract

Purpose

Total shoulder arthroplasty (TSA) has become a successful treatment option for degenerative shoulder disease. With the increasing incidence in primary TSA procedures during the last decades, strategies to improve implant longevity become more relevant. Implant failure is mainly associated with mechanical or biological causes. Chronic inflammation as a response to wear particle exposure is regarded as a main biological mechanism leading to implant failure. Metal ions released by fretting and corrosion at modular taper connections of orthopedic implants can cause cell-mediated hypersensitivity reactions and might lead to aseptic loosening. Modularity is also commonly used in total shoulder replacement. However, little is known about metal ion exposure in patients following TSA. The objective of this study was to determine in-vivo blood metal ion levels in patients after TSA and to compare blood metal ion levels to control subjects without metal implants.

Methods

A total of 19 patients with anatomical total shoulder prosthesis (TSA group) and 20 patients with reverse total shoulder prosthesis (RSA group) who underwent unilateral total shoulder replacement at our hospital between March 2011 and December 2014 with no other metal implant or history of environmental metal ion exposure were recruited for analysis of blood metal ion concentrations of cobalt (Co), chromium (Cr) and titanium (Ti) at a mean follow-up period of 2.3 years (0.7–4.3). For comparison of metal ion concentrations blood samples were obtained in a healthy control group of 23 subjects without metal implants. Ethical approval and informed consent of each patient were obtained for this study.

Results

Median cobalt ion levels were 0.14µg/l (range 0.03–0.48) in the TSA group, 0.18 µg/l (0.10–0.66) in the RSA group and 0.11µg/l (0.03–0.19) in the control goup. Median chromium ion levels were 0.34µg/l (0.09–1.26) in the TSA group, 0.48µg/l (0.17–2.41) in the RSA group and 0.14µg/l (0.04–0.99) in the control goup. Median titanium ion levels were 0.86µg/l (0.10–1.64) in the TSA group, 1.31µg/l (0.75–4.52) in the RSA group and 0.62µg/l (0.32–2.14) in the control goup. There was a statistically significant difference in chromium and titanium ion concentrations between both study groups and the control group (see figure 1–3).

Conclusion

Patients with unilateral total shoulder replacement demonstrated elevated blood metal ion concentrations. Median blood metal ion levels were higher in the RSA group compared to the TSA group, which could be attributable to the modularity of the reverse total shoulder system. However, overall metal ion levels were relatively low compared to those seen in patients with metal-on-metal total hip replacements. The role of local metal ion exposure in the development of aseptic loosening or hypersensitivity reactions associated with total shoulder arthroplasty should be further investigated.

For figures/tables, please contact authors directly.