3D-PRINTED PATIENT-SPECIFIC INSTRUMENTATION IN ORTHOPAEDIC ONCOLOGY: IMPACT OF CUT SLOT DEPTH ON ACCURACY

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 3.



Abstract

Introduction

Clear operative oncological margins are the main target in malignant bone tumour resections. Novel techniques like patient specific instruments (PSIs) are becoming more popular in orthopaedic oncology surgeries and arthroplasty in general with studies suggesting improved accuracy and reduced operating time using PSIs compared to conventional techniques and computer assisted surgery. Improved accuracy would allow preservation of more natural bone of patients with smaller tumour margin.

Novel low-cost technology improving accuracy of surgical cuts, would facilitate highly delicate surgeries such as Joint Preserving Surgery (JPS) that improves quality of life for patients by preserving the tibial plateau and muscle attachments around the knee whilst removing bone tumours with adequate tumour margins. There are no universal guidelines on PSI designs and there are no studies showing how specific design of PSIs would affect accuracy of the surgical cuts.

We hypothesised if an increased depth of the cutting slot guide for sawblades on the PSI would improve accuracy of cuts.

Methods

A pilot drybone experiment was set up, testing 3 different designs of a PSI with changing cutting slot depth, simulating removal of a tumour on the proximal tibia (figure 1)

A handheld 3D scanner (Artec Spider, Luxembourg) was used to scan tibia drybones and Computer Aided Design (CAD) software was used to simulate osteosarcoma position and plan intentioned cuts (figure 1). PSI were designed accordingly to allow sufficient tumour. The only change for the 3 designs is the cutting slot depth (10mm, 15mm & 20mm). 7 orthopaedic surgeons were recruited to participate and perform JPS on the drybones using each design 2 times. Each fragment was then scanned with the 3D scanner and were then matched onto the reference tibia with customized software to calculate how each cut (inferior-superior-vertical) deviated from plan in millimetres and degrees (figure 3). In order to tackle PSI placement error, a dedicated 3D-printed mould was used.

Results

Comparing actual cuts to planned cuts, changing the height of the cutting slot guide on the designed PSI did not deviate accuracy enough to interfere with a tumour resection margin set to maximum 10mm. We have obtained very accurate cuts with the mean deviations(error) for the 3 different designs were: [10mm slot: 0.76±0.52mm, 2.37±1.26°], [15mm slot: 0.43±0.40mm, 1.89±1.04°] and [20mm: 0.74±0.65mm, 2.40±1.78°] respectively, with no significant difference between mean error for each design overall, but the inferior cuts deviation in mm did show to be more precise with 15mm cutting slot (p<0.05) (figure 2).

Discussion

Simulating a cut to resect an osteosarcoma, none of the proposed designs introduced error that would interfere with the tumour margin set. Though 15mm showed increased precision on only one parameter, we concluded that 10mm cutting slot would be sufficient for the accuracy needed for this specific surgical intervention. Future work would include comparing PSI slot depth with position of knee implants after arthroplasty, and how optimisation of other design parameters of PSIs can continue to improve accuracy of orthopaedic surgery and allow increase of bone and joint preservation.