Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

FRETTING AND CORROSION DAMAGE IN SLEEVED CERAMIC HEADS: A RETRIEVAL STUDY

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 2.



Abstract

Introduction

During revision surgery with a well-fixed stem, a titanium sleeve can be used in conjunction with a ceramic head to achieve better stress distribution across the taper surface. Previous studies have observed that the use of a ceramic head can mitigate the extent of corrosion damage at the taper. Moreover, in vitro testing suggests that corrosion is not a concern in sleeved ceramic heads [1]; however, little is known about the in vivo fretting corrosion of the sleeves. The purpose of this study was to investigate fretting corrosion in sleeved ceramic heads.

Materials and Methods

Thirty sleeved ceramic heads (Biolox Option: CeramTec) were collected during revision surgery as part of a multi-center retrieval program. The sleeves were used in conjunction with a zirconia-toughened alumina femoral head. The femoral heads and sleeves were implanted between 0.0 and 3.25 years (0.8±0.9, Figure 1). The implants were revised predominantly for instability (n=14), infection (n=7), and loosening (n=5). Fifty percent of the retrievals were implanted during a primary surgery, while 50% had a history of a prior revision surgery. Fretting corrosion was scored using a previously described 4-point, semi-quantitative scoring system proposed by Higgs [2].

Results

Among the sleeved ceramic heads, mild-to-moderate fretting corrosion scores (Score = 2–3) were observed in 96% of internal tapers, 26% of external tapers, and 82% of the stems. On the internal taper surface, 5 sleeves had moderate fretting corrosion data (Score = 3, Figure 2). None of sleeves had severe (Score = 4) at any taper surface. Fretting corrosion scores were higher at the internal taper surface than the external taper. Implantation time was the main predictor of increased fretting-corrosion of the external sleeve tapers.

Discussion

For the sleeved ceramic heads, we found that fretting corrosion can occur in these components, particularly on the internal surface of the sleeve. However, the fretting corrosion scores were predominantly mild, and lower than fretting scores of CoCr heads in metal on polyethylene bearings. Because the sleeves are Ti alloy, the corrosion products are considered to be less cytotoxic than Co and Cr. The primary limitation to this study is the short-term follow-up of these retrievals. As the fretting corrosion process is often associated with in vivo duration, future studies with longer-term implants are necessary to elucidate the long-term performance of these devices.


*Email: