ACCURACY OF GLENOID VERSION AND INCLINATION: A COMPARISON OF SURGICARE AND BLUEPRINT SURGICAL PLANNING SOFTWARE PLATFORMS

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 2.



Abstract

Background

Virtual planning of shoulder arthroplasty has gained recent popularity. Combined with patients specific instrumentation, several systems have been developed that allow the surgeon to accurately appreciate and correct glenoid deformities in version and inclination. While each virtual software platform utilizes a consistent algorithm for calculating these measurements, it is imperative for the surgeon to recognize any differences that may exist amongst software platforms and characterize any variability.

Methods

A case-control study of all CT scans of patients previously pre-operatively planned using MatchPoint SurgiCase® software were uploaded into the BluePrint software. The cohort represents surgical planning for total shoulder arthroplasty and reverse shoulder arthroplasty with varying degrees of glenoid deformity. Glenoid version and inclination will be recorded for each CT scan using both software platforms.

Results

A total of 38 patient CT scans previously planned using MatchPoint Surgicase® software were uploaded into the BluePrint software. The mean difference for glenoid version between the two software programs was 2.497° (±1.724°) with no significant differences in measured glenoid version readings between BluePrint and SurgiCase software (p=0.8127). No significant differences were seen in the measured glenoid inclination between the two software programs (p=0.733), with a mean difference for glenoid inclination between the two software programs at 5.150° ± 3.733° (figure 1). A Bland-Altman plot determined the 95% limits of agreement between the two programs at −5.879 to 6.116 degrees of glenoid version and −12.05 to 12.75 degrees of glenoid inclination. There was a significant statistical agreement between the two software programs measuring glenoid version and inclination in relation to glenoid wear position for the centered (p=0.004), posterior (p<0.001, p=0.003), posterior-superior (p<0.001, p<0.001), and superior (p=0.027, p=0.034) positions, respectively.

Conclusions

Both BluePrint and SurgiCase software platforms yield similar measurements for glenoid version and glenoid inclination. In the setting of glenoid wear in the posterior, posterior-superior or superior position, measurements of between two surgical platforms are in agreement.