Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

A NOVEL TAPERED WEDGE STEM: MULTICENTRE CLINICAL OUTCOMES STUDY

The International Society for Technology in Arthroplasty (ISTA), 29th Annual Congress, October 2016. PART 1.



Abstract

Introduction

Cementless tapered wedge stems have shown excellent results over the last decade. Distal potting with inadequate proximal fit, as well as failure to achieve biologic fixation has led to thigh pain, loosening and implant failure. To support a variety of patient morphologies a novel tapered wedge stem was designed with reduced distal morphology, maximizing the proximal contact of the grit blasted surface. The objective of the study was to analyze the clinical outcomes of this stem design.

Methods

Three hundred and nineteen patients enrolled into prospective, post-market multicenter studies received a novel tapered wedge stem. Clinical and patient-reported outcomes including the Harris Hip Score (HHS), Lower Extremity Activity Scale (LEAS), Short Form 12 (SF12), and Euroqol 5D Score (EQ-5D) were evaluated preoperative through two years postoperative.

Results

Demographics of the study population include 53.3% male patients with a mean age of 62.4 ± 9.3 years and BMI of 29.8 ± 4.5. To date, there has been one incidence of thigh pain (0.3%) and a 0.63% aseptic revision rate. Kaplan-Meier survivorship analysis for the population at two years postoperative estimated 97.87% with 95% CI (93.98% – 99.25%) survivorship to all cause revision (Figure 1). Statistically significant improvements in pain, function, and quality of life were seen initially at six weeks postoperative and continued through one year in all of the clinical and patient reported outcome measures (Figure 2).

Conclusion

A novel tapered wedge stem was designed to maximize proximal fit medially and laterally across a variety of patient morphologies. Radiographic studies have shown the stem design has significantly better canal fit compared to older tapered wedge designs. The current study exhibits low incidence of thigh pain and revisions, with improvements in function, pain and quality of life after hip arthroplasty with this tapered wedge stem.


*Email: