Abstract
In primary total hip replacements there are numerous options available for providing hip stability in difficult situations (i.e. Down's syndrome, Parkinson's disease). We have considered constrained liners in some of these cases.
However, in the revision situation in general and in revision for recurrent dislocation situation specifically it is important to have all options available including tripolar constrained liners in order to optimise the potential for hip stability as well as function of the arthroplasty. Even with the newer options available dislocation rates of higher than 10–15% have been reported following revision surgery at institutions where high volumes of revision surgery are performed. Because of the deficient abductors, other soft tissue laxity and the requirement for large diameter cups revision cases will always have more potential for dislocation. In these situations in the lower demand patient, constraint has provided excellent success in terms of preventing dislocation and maintaining implant construct fixation to bone at intermediate- term follow-up. Hence in these situations tripolar constrained liners remains the option we utilise. We are also confident in using this device in cases with instability or laxity where there is a secure well- positioned acetabular shell. We cement a dual mobility constrained liner in these situations using the technique described below.
Present indication for tripolar constrained liners: low demand patient, large outer diameter cups, instability with well-fixed shells that are adequately positioned, abductor muscle deficiency or soft tissue laxity, multiple operations for instability
Technique of cementing liner into shell: score acetabular shell if no holes, score liner in spider web configuration, all one or two millimeters of cement mantle
Results
Constrained Dual Mobility Liner
For Dislocation: 56 Hips, 10 yr average f/u, 7% failure of device, 5% femoral loosening, 4% acetabular loosening
For Difficult Revisions:101 hips, 10 yr average f/u, 6% failure of device, 4% femoral loosening, 4% acetabular loosening
Cementing Liner into Shell: 31 hips, 3.6 yr average f/u (2–10 years), 2 of 31 failures