Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

TRUNNION CORROSION: IS IT REALLY A PROBLEM?

The Current Concepts in Joint Replacement (CCJR) Spring Meeting, Las Vegas, May 2017.



Abstract

Uncemented metal-on-polyethylene total hip arthroplasties (THAs) have had a modular cobalt-chrome alloy head since their introduction in the early 1980's. Retrieval analysis studies and case reports in the early 1990's first reported corrosion between the femoral stem trunnion (usually titanium alloy) and cobalt-chrome alloy femoral head. However, then this condition seemed to disappear for about two decades? There are now numerous recent case series of this problem after metal-on-polyethylene THA, with a single taper or dual taper modular femoral component. Metal ion elevation, corrosion debris, and effusion are caused by mechanically assisted crevice corrosion (MACC). These patients present with diffuse hip pain, simulating trochanteric bursitis, iliopsoas tendinitis, or even deep infection. Trunnion corrosion, with adverse local tissue reaction, is a diagnosis of exclusion, after infection, loosening, or fracture. The initial lab tests recommended are: ESR, CRP, and serum cobalt and chromium ions. With a metal-on-polyethylene THA, a cobalt level > 1ppb is abnormal. Plain radiographs are usually negative, but may show calcar osteolysis or acetabular erosion or cyst. MARS MRI may be the best imaging study to confirm the diagnosis. Hip aspiration for culture and cell-count may be necessary. The operative treatment is empiric, with debridement, and head exchange with a ceramic head-titanium sleeve (or oxidised zirconium head) placed on the cleaned trunnion. The femoral component may have to be removed if there is “whole trunnion failure”. This usually relieves the symptoms, but the complication rate of this procedure may be high.