DYNAMIC ANALYSIS OF DISLOCATION-ASSOCIATED FACTORS IN TOTAL HIP REPLACEMENT BY HARDWARE-IN-THE-LOOP SIMULATION

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress. PART 2.



Abstract

Introduction

The purpose of this study was to experimentally evaluate impingement and dislocation of total hip replacements while performing dynamic movements under physiological-like conditions. Therefore, a hardware-in-the-loop setup has been developed, in which a physical hip prosthesis actuated by an industrial robot interacts with an in situ-like environment mimicked by a musculoskeletal multibody simulation-model of the lower extremity.

Methods

The multibody model of the musculoskeletal system comprised rigid bone segments of the lower right extremity, which were mutually linked by ideal joints, and a trunk. All bone geometries were reconstructed from a computed tomography set preserving anatomical landmarks. Inertia properties were identified based on anthropometric data and by correlating bone density to Hounsfield units. Relevant muscles were modeled as Hill-type elements, passive forces due to capsular tissue have been neglected. Motion data were captured from a healthy subject performing dislocation-associated movements and were fed to the musculoskeletal multibody model. Subsequently, the robot moved and loaded a commercially available total hip prosthesis and closed the loop by feeding the physical contact information back to the simulation model. In this manner, a comprehensive parameter study analyzing the impact of implant position and design, joint loading, soft tissue damage and bone resection was implemented.

Results

The parameter study revealed a generally high dislocation risk for the seating-to-rising with adduction scenarios. Improper implant positioning or design could be compensated by adjusting prosthesis components correspondingly. Gluteal insufficiency or lower joint loading did not result in higher impingement or dislocation risk. However, severe malfunction of the artificial joint was found for proximal bone resection.

Discussion

Previous testing setups ignored the impact of active muscles or relied on simplified contact mechanics. Herein, total hip replacement stability has been investigated experimentally by using a hardware-in-the-loop simulation. Thereby, several influencing factors such as implant position and design as well as soft tissue insufficiency and imbalance could be systematically evaluated with the goal to enhance joint stability.