Abstract
Aim
The aim of this study is to evaluate the effect of three-dimensional (3D) simulation with 3D planning software ZedKnee® (ZK) in total knee arthroplasty (TKA).
Materials and methods
The participants in this study were all TKA patients whose operations were simulated by using ZK. The alignment of all components was evaluated with the ZK valuation software in postoperative computer tomography. Thirty patients (43 knees) met the inclusion criteria. 6 patients were male and 24 patients were female. The mean age of the 30 patients was 72 years old. Diagnoses for surgery were: osteoarthritis- 40 knees, rheumatoid arthritis- 2 knees and osteonecrosis- 1 knee. TKA was performed using the measured resection technique. The distal femur axis where the intramedullary rod would be inserted was drawn manually on the 3D image. Then, the angle between the distal femoral axis and the mechanical axis was measured. The rotational angles of the femoral components were determined from the automatically calculated angle between the posterior condylar axis and the surgical epicondylar axis (SEA) by using ZK. The ZK data used during the operation was the posterior condylar angle, the angle between the distal femoral axis and the mechanical axis and implant size.
Results
The angle in coronal plane between the 3D mechanical axis and the distal femoral axis in preoperative planning ranged between 3 degrees and 11 degrees, mean 6.7 (SD 2.2) degrees. The postoperative femoral component alignment was on average 0.7 (SD 1.3) degrees in varus. Outlier of more than 3 degrees in coronal alignment was recognized in 3 cases (7%). The mean posterior condylar angle in preoperative planning was 3.8 (SD 1) degrees. The postoperative femoral component alignment was on average 1.5 (SD 1.6) degrees in external rotation to surgical epicondylar axis. Outlier of more than 3 degrees in rotational alignment was recognized in 6 cases (14%). The concordance rate between the preoperative planning size and the intraoperative selective size was 91%.
Discussion
Some errors may be observed in the preoperative TKA X-ray planning, because of the rotational position of the femur while having the X-ray taken or angle of the X-ray beam. Kanekasu et al reported the measurement of the condylar twist angle during the X-ray and it was relatively correct compared with the measurement during CT. Max 1.9 degrees error occurred in the measurements using X-rays. It appeared that preoperative planning using CTs was more accurate than using X-rays.
Conclusion
Femoral components with 3D simulation using ZK were fixed perpendicularly against the mechanical axis and parallel to the surgical epicondylar axis with high accuracy. We considered that the ZK 3D simulation in TKA is useful for the accurate alignment of femoral components.