header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

IMPROVING DIAGNOSTIC ACCURACY OF HIP DYSPLASIA MEASURES IN 2D ULTRASOUND SCANS OF INFANTS TO GUIDE DECISIONS REGARDING NEED FOR SURGERY

International Society for Computer Assisted Orthopaedic Surgery (CAOS) - 15th Annual Meeting



Abstract

Ultrasound (US) imaging is recommended for early detection of Developmental Dysplasia of the Hip (DDH) to guide decisions about possible surgical treatment. However, a number of studies have raised concerns over the efficacy of US in early diagnosis. The main limitation of US-based diagnosis is sub-standard reliability of the primary dysplasia metric measurements: namely, the alpha and beta angles. In this study, we have proposed a novel and automatic method to extract dysplasia metrics from 2D US, which we hope will improve the overall reliability of US-based DDH measurements by removing error due to subjective measurements. We hypothesise that improvements in reliability of dysplasia metric measurements will reduce the chances of missed early-diagnosis, and therefore reduce the need for later complex surgical treatments.

We evaluated performance of the algorithm on 4 infants diagnosed with US scans for DDH. The typical runtime of our algorithm is less than 1 second for an US image. We found a 6° bias between manual and automatic measurements, with automatic measurements tending to be lower in value; the standard deviation in the discrepancy values was also relatively high at 7°. This suggests that there is considerable variability in the angle estimation process, which is typically done manually, which supports our contention that further work needs to be done to establish an accurate and repeatable measurement technique. Further, we found agreements in the Graf-classification types in six out of seven sessions. For the one patient where there was a discrepancy in classification, later US sessions suggest the manual technique possibly missed the opportunity for early detection, in contrast to the automatic method which classified the patient as having evidence of dysplasia. Thus, such an automatic method may improve the reliability of current US-based DDH diagnosis techniques. The primary limitation of this study is that we have done strictly an intra-image discrepancy analysis and have not compared the results with what could be considered a ‘gold standard’ reference. In future work, we plan to assess these indices on 3D images of the hip and assess the accuracy of proposed 2D and 3D-based automatic index calculation techniques against a 3D reference model.