USEFULNESS OF CT-BASED PRE-OPERATIVE PLANNING FOR SEVERE TKAs

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 4.



Abstract

Introduction

Computed tomography (CT) based preoperative planning provides useful information for severe TKA and revision TKA cases, such as the amount of augmentation, length of stem extension and component alignment, to achieve correct alignment and joint line. In this study, we evaluated TKA alignment performed with CT preoperative planning.

Materials and Methods

7 primary TKAs for severe deformity and 3 revision TKAs were included. CT preoperative planning was performed with JIGEN (LEXI, Japan). Constrained condylar prosthesis (LCCK, Zimmer) were used in all case. For femoral component, axial alignment was decided by controlled IM rod insertion to femoral canal. Rotational alignment was decided according to anterior cortex that usually was not compromised. For tibial component, axial alignment was set to perpendicular to tibial mechanical axis. Coverage and joint line level were carefully decided. The amount of bone resection of bilateral distal and posterior femoral condyle and proximal tibia was measured, respectively. Stem extension length and offset were selected according to components position and canal filling. Amount of augmentation was also estimated bilateral distal and posterior femoral condyle, respectively. Postoperative component alignment was evaluated three-dimensionally with Knee-CAS (LEXI, Japan).

Results

All femoral and tibial components were implanted within 5°in coronal and sagittal plane. All knees showed mechanical alignment within 5 degree from neutral. One of 10 TKAs needed femoral component size down, and two of 20 stems needed size change.