Abstract
Purpose
A Trabecular Metal Modular Acetabular System (Zimmer, Warsaw, Indiana, USA) is a peripheral rim expansion (elliptical) cup, i.e. a non-hemispherical cup. Radiologically a non-hemispherical cup may be deferent from other conventional hemispherical cups. We reviewed radiological findings of a Trabecular Metal Modular Acetabular System chronologically.
Methods
Twenty six patients with osteoarthritis underwent primary total hip arthroplasty (THA) using a Trabecular Metal Modular Acetabular System from 2011 to April 2013. Twenty five patients (follow-up rate: 96.2%) 31 hips could be followed-up over a year were registered. In common, the diameter of every femoral head was 32 mm. We planned the acetabular cup inclination angle to be 45-degree, the cup coverage with host-bone (cup-CE angle) to be over 10-degree, and high hip center was allowed up to 20mm. In case of the cup-CE angle under 10-degree, an acetabular cup was placed medially using Dorr's medial protrusio technique. We established the medial protrusion angle indicating the degree of medial protrusion of an acetabular cup over the pelvic internal wall. The medial protrusion angle was defined by the center point of THA (C) and the 2 cross-points (X1, X2) which the outline of an acetabular cup crosses the Kohler's line (Figure 1). The cup anteversion angle was measured by the method of Lewinnek, and the cup fixation was evaluated according to the Tompkin's classification.
Results
The average follow-up period was 1 year and 3 months (1y1m to 2y8m). The mean diameter of the cup was 54 (48 to 56) mm. Seven high-hip center joints were recognized (2 to 11 mm). The average of cup inclination angle was 42 (32 to 52) degree, of cup anteversion angle was 14 (5 to 36) degree, and of cup CE angle was 25 (−14 to 45) degree. Dorr's medial protrusio technique was necessary in 18 hips. In these 18 hips, the average of medial protrusion angle was 57 (24 to 70) degree. In 4 hips of cup-CE angle less than 10 degree, acetabular bulky bone graft was added. All 31 hips showed the stable fixation, even in 18 hips undergone medial protrusio technique. There was none of hips with migration and/or rotation of an acetabular cup. Radiolucent zone was found in the zone-C of 8 hips. The width of radiolucent zone of all 8 hips was less than 2mm. In these 8 hips, medial protrusio technique was done in 5 hips, and high hip center was found in 3 hips. The radiolucency appeared at postoperatively 2–3 months and disappeared by postoperatively 12 months.
Conclusions
All hips showed rigid fixation of a Trabecular Metal Modular Acetabular System in short-term observation. Even in the hips performed Dorr's medial protrusio technique, a Trabecular Metal Modular Acetabular System reached the stable fixation. Radiolucent zone was found transiently in the zone-C of 8 hips (25.8%) and disappeared by postoperatively 12 months. However our series was small and the observation period was short, our results implied that the fixation of a Trabecular Metal Modular Acetabular System was not affected adversely from Dorr's medial protrusio technique.