General Orthopaedics

HIGH INCIDENCE OF STEM LOOSENING IN ASSOCIATION WITH PERIPROSTHETIC FEMORAL FRACTURES IN PREVIOUSLY WELL-FIXED CEMENTLESS GRIT-BLASTED TAPERED-WEDGE STEMS

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 3.



Abstract

Introduction

Cementless grit-blasted tapered-wedge titanium femoral stems are being used with increasing frequency in hip arthroplasty because of excellent long-term outcomes. However, periprosthetic femur fracture is a potentially worrisome phenomenon in these types of femoral stems. The aim of this study is to report the incidence of stem loosening in association with periprosthetic femur fractures following hip arthroplasty using cementless grit-blasted tapered-wedge stems.

Materials & Methods

A total of 36 Vancouver Type B1 and B2 periprosthetic femur fractures following either hemiarthroplasty or total hip arthroplasty using cementless grit-blasted tapered-wedge titanium femoral stems (GB group) were identified from a retrospective review of the medical records at three participating academic institutions. The control group consisted of 21 Vancouver Type B1 and B2 periprosthetic femur fractures following either hemiarthroplasty or total hip arthroplasty using cementless proximal porous-coated femoral stems (PC group) at the same institutions during the same period of the study. All femoral stems included in this study had been a well-fixed state before the occurrence of periprosthetic femur fractures. All patients in both groups were treated surgically with either open reduction and internal fixation or femoral stem revision. Femoral stem stability was assessed by preoperative radiographs and was confirmed by intraoperative scrutinization. The incidence of stem loosening was compared between the groups.

Results

There was no significant difference between the groups with respect to demographic data including age, gender, body mass index, primary diagnosis, Dorr types of proximal femur, and time to fracture. All fractures occurred from low-energy mechanisms. Mean age at the time of hip arthroplasty was 54.5 years in the GB group and 57.0 years in the PC group. Mean time interval between hip arthroplasty and periprosthetic fracture was 49.6 months in the GB group and 44.4 months in the PC group. At the time of the last follow-up, 29 (80.6%) of 36 fractures was Vancouver B2 in the GB group, whereas only 3 (14.3%) of 21 fractures was Vancouver B2 in the PC group (P <0.001).

Conclusions

High incidence of stem loosening was developed in association with periprosthetic femur fractures in previously well-fixed cementless grit-blasted tapered-wedge femoral stems in our population. We believe that this is an underreported phenomenon of these types of stem design.