Abstract

The purpose of this study was to determine the effect of sectioning the relevant soft tissues and a TKA on the medial and lateral knee joint gap.

Twelve intact lower extremity cadaveric specimens (mean (SD) age 76.5 (11.6) years) were tested. A custom designed knee tensioner was developed that allowed the separate application of forces to the medial and lateral components of the knee. The distance between the bottom of the load cell and the top of a compression rod was measured with digital calipers (precision = 0.1mm). Loads of 100N and 200N were then applied to each compartment and the resulting displacement was measured. The two loads were applied to the knee in the following conditions: i) All soft tissues intact; ii) an arthrotomy; iii) ACL sectioned; iv) PCL sectioned; v) release of the mid-coronal tissues; and vi) TKA. Finally, tensions were applied for all conditions from 90° to 0° of knee flexion in 30° increments.

There was a significant effect of soft tissue release on the magnitude of the gap at the 100N load application, such that there was an increase in the when the mid-coronal MCL release was performed compared to the intact (2.2mm) and arthrotomy (1.75mm) conditions. With respect to the 200N load application there was a statistically significant tissue release effect, where differences were detected between the mid-coronal MCL release and intact (3.04mm) and arthrotomy conditions (2.31mm). At the 100N load there was a significance increase in the gap compared to the intact knee. There was also a significant condition by knee angle interaction where the gap was approximately 4mm larger following the TKA compared to the intact condition when the knee was flexed at 90°. Furthermore, there was a statistically significant 4.8mm and 3.8mm difference between 90° and 0° and 60° and 0° of knee flexion respectively, for the TKA condition only. At the 200N load application the gap width increased significantly by 2.5mm following the TKA. Finally, there was a significant condition by knee angle interaction where the change in gap width increased significantly from the intact (7.54mm) to the TKA condition (13.88mm) at 90° of knee flexion. There was a statistically significant difference in the TKA condition between 60° and 0° of knee flexion.

Releasing the soft tissues increases the gap between the tibia and femur, when compared to the intact condition, with significance occurring only following the mid-coronal release. Furthermore, the TKA did not return the knee to its intact state as was evident by the significant difference between the TKA and intact conditions. This suggests that the resulting kinematics may not accurately match those pre-surgery resulting in un-physiological motion patterns and the possibility of early failure and revision.