Modulus femoral prosthesis is a modular cementless femoral system which consists of 5 degree tapered conical stem made of a titanium alloy with 8 fins of 1mm and modular neck. Modular neck enables to control any ideal stem anteversion as a surgeon prefers. This system is considered to be useful in severe hip deformity, for example developmental dysplasia of the hip (DDH). In this study, clinical and radiographic outcomes of the Modulus femoral prosthesis were evaluated at a mean of 3.6 (2–6) years postoperatively.
We assessed the results of 193 primary total hip arthroplasty using a Modulus femoral prosthesis in 169 patients (15 males, 154 females) undergoing surgery between September 2007 and December 2011. The mean age at the time of surgery was 65.6 (31–86) years old. The diagnoses were osteoarthritis (OA) in 178 hips (including 167 hips of DDH), rapidly destructive coxopathy (RDC) in 6 hips, rheumatoid arthritis (RA) in 6 hips, osteonecrosis in 2 hips, and subchondral insufficiency fracture in one hip. Clinical outcomes were assessed using Japan Orthopedic Association (JOA) hip scores and complications. Radiographic assessments were including stem alignment, bone on-growth, cortical hypertrophy, stress shielding and stem subsidence.
43.8 points of the preoperative mean JOA score was significantly improved to 93.1 points postoperatively. In one case intraoperative femoral fracture was occurred. One dislocation had occurred and thigh pain was observed in one hip. No revision surgery was required. In 192 hips of 193 hips (99.5%), stem was implanted in neutral position (within ±2 degrees). Bone on-growth was observed in all cases (94.3% in zone 3; 73.1% in zone 5; 30.6% in zone 2; 22.3% in zone 6). Cortical hypertrophy was observed in 66 hips (34.2%) at zone 3 and 5. Reduction of bone density due to stress shielding was observed (1st degree was 58.5%; 2nd degree was 29.5%; 3rd degree was 11.9%; 4th degree was 0%). In 22 cases (11.4%), more than 2mm of stem subsidence was observed, however the subsidence was stopped within 6 months in all cases.
Modulus femoral prosthesis showed good clinical results and radiographic findings up to 6 years postoperatively.