THE STRESS DISTRIBUTION IN THE FEMORAL MEDULLARY CAVITY FORM AND SURROUNDING OF HIP STEMS: FINITE ELEMENT ANALYSES

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress, 2015. PART 4.



Abstract

Introduction

On the basis of a proposal by Noble, the marrow cavity form can be classified into three categories: normal, champagne-fluted and stovepipe. In the present study, three typical finite element femoral models were created using CT data based on Noble's three categories. The purpose was to identify the relationship of stress distribution of the surrounding areas between femoral bone marrow cavity form and hip stems. The results shed light on whether the distribution of the high-stress area reflects the stem design concept. In order to improve the results of THA, researchers need to consider the instability of a stem design based on the stress distributioin and give feedback on future stem selection.

Methods

As analyzing object, we selected SL-PLUS and BiCONTACT stems. To develop finite element models, two parts (cortical bone and stem) were constructed using four-node tetrahedral elements. The model consisted of about 60,000 elements. The material characteristics were defined by the combination of mass density, elastic coefficient, and Poisson's ratio. Concerning the analysis system, HP Z800 Workstation was used as hardware and LS-DYNA Ver. 971 as software. The distal end of the femur was constrained in all directions. On the basis of ISO 7206 Part 4,8 that specifies a method of endurance testing for joint prostheses, the stem was tilted 10°, and a 1500 N resultant force in the area around the hip joint was applied to the head at an angle of 25° with the long axis. Automatic contact with a consideration of slip was used.

Result

The maximum stress on femur implanted a SL-PLUS with marrow cavity form of normal, champagne-fluted and stovepipe were shown to be 90MPa, 90MPa and 45MPa. The maximum stress on a BiCONTACT with marrow cavity form of normal, champagne-fluted and stovepipe were shown to be 45MPa, 90MPa and 15MPa.

Discussion

The design concept for aZweymüller-type stem can distribute load across a wide range of cortical bone from the middle position to the distal femur. It is determined using this concept that a wide range of stress was absorbed at the middle position and distal femur in the champagne-fluted and normal cases. On the other hand, the contact pressure zone of stovepipe could not meet the expected level at the distal femur. The method of this research involves controlling the stress conditions within the stem design. At this point, it is considered possible for the stability of various stem designs to be predicted and the stability to be assessed positively. On the basis of Noble's categories, three types of finite element model were made, and stress distribution measurement and finite element analyses were performed. The results indicate that Zweymüller stem has clinical validity for securing force in the champagne-fluted and stovepipe types from the stress distribution.