header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

ADVANTAGES AND DRAWBACKS OF A NEW VOLAR PLATE MADE OF CARBON-FIBRE-REINFORCED POLYETHERETHERKETONE FOR DISTAL RADIUS FRACTURE

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress, 2015. PART 4.



Abstract

Background

Implants based on the polyetheretherketon (PEEK) polymer have been developed in the last decade as an alternative to conventional metallic devices. PEEK devices may provide several advantages over the use of conventional orthopedic materials, including the lack of metal allergies, radiolucency, low artifacts on magnetic resonance imaging scans and the possibility of tailoring mechanical properties. The purpose of this study was to evaluate the clinical results at mean 24-month follow-up using a new plate made of carbon-fiber-reinforced polyetheretherketon (CFR-PEEK) for the treatment of distal radius fractures.

Materials and methods

We performed a prospective study including all patients who were treated for unstable distal radius fracture with a CFR-PEEK volar fixed angle plate. We included 70 consecutive fractures of AO types B and C that remained displaced after an initial attempt at reduction. The fractures were classified according to the AO classification: 35 fractures were type C1, 13 were type C2, 6 were type C3, 5 were type B1 and 11 were type B2.

Results

All fractures healed, and radiographic union was observed at an average of 6 weeks. The final Disabilities of Arm, Shoulder and Hand score was 5.2 points. The average grip strength, expressed as a percentage of the contralateral limb, was 94 %. Three cases of hardware breakage were reported. Two cases were due to intraoperative plate rupture caused by the attempt to achieve the reduction of the fracture in 1 case and while inserting a distal screw in the other case. In the last case hardware breakage was caused by a fall on the injuried arm 1 week after surgery. No cases of loss of the surgically achieved fracture reduction were documented. Hardware removal was performed in 3 cases, for the occurrence of extensor tenosynovitis in 2 patients and tenosynovitis of flexor pollicis longus in 1 case.

Conclusion

The major advantage of CFR-PEEK plate is its radiolucency. This characteristic allows direct visualization of osseous callus formation, allowing monitoring of the healing of the fracture, thereby improving clinical assessment and accuracy. Therefore, specific indications for this new radiolucent plate can be represented by fractures with significant metaphyseal comminution and in cases of nascent malunion where a distal radius osteotomy with bone grafting is usually performed to correct the wrong angle. At early follow-up this device showed good clinical results and allowed maintenance of reduction in complex, AO fractures. The occurrence of tendon complications related to this implant was similar to that reported in literature for the other new-generation plates. However, attention should be payed when stressing the plate to achieve the desired fracture reduction to avoid hardware failure.


*Email: