Different bearing surfaces, including alumina-on-alumina have been used to avoid osteolysis.
We ask if the use of modern ceramics matched with metal-backed sockets improves clinical results avoiding aseptic loosening, osteolysis and late dislocations.
We prospectively analysed 315 Cerafit cups in two different generations: First generation implanted between 1999 and 2005 (124 cups) and second generation (191 cups) all implanted with a 5-year minimum follow-up, and associated with a Multicone-HAP stem using alumina-on-alumina bearings. Patients' mean age was 48.7+13.6 years.
There were three dislocations within the first postoperative month and one early recurrent dislocation requiring reoperation using a double mobility cup. There were four intraoperative fractures, two in rheumatoid juvenile arthritis cases solved with a cerclage, and one postoperative fracture solved using a femoral long-stem. No fractures were seen in the alumina components in this series. Also, there have been no infections. There were 4 patients reporting non-reproducible squeaking. Survivorship analysis for cup loosening for any cause was 93% for group 1 (11 cups) and 99.2% for group 2 (1 cup in a patient with severe developmental dysplasia). The cumulative probability of not having a revision of one or both components for any cause was 96.9% (CI 95%:9.47–99.1%), All unrevised cases showed good clinical and radiographic results at the end of follow-up. No stem loosening, osteolysis nor stress shielding were found in any case by the end of follow-up. Changes in linear femoral head penetration were not seen in any hip.
These data suggest that Cerafit alumina-on-alumina prostheses show excellent results after fiften years. Cup loosening was more frequent in the cups of first generation, but not currently. Continued follow-up will be required to determine if reduction in wear between the alumina-on alumina bearings results in less osteolysis and loosening.