General Orthopaedics

IN VIVO KINEMATICS OF A MEDIALLY CONFORMING AND ROTATIONALLY UNCONSTRAINED TKA DESIGN

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 1.



Abstract

Introduction

Total knee arthroplasty (TKA) designs evolve as evidence accumulates on natural and prosthetic knee function. TKA designs based upon a medially conforming tibiofemoral articulation seek to reproduce essential aspects of normal knee stability and have enjoyed good clinical success and high patient satisfaction for over two decades. Fluoroscopic kinematic studies on several medially conforming knee designs show extremely stable knee function, but very small ranges of tibial axial rotation compared to healthy knees. The GMK Sphere TKA is a recent evolution in medially-conforming TKA designs that adopts a sagittally unconstrained lateral tibiofemoral articulation to allow more natural tibial rotation. This study was conducted to quantify motions in knees with this prosthesis to address two questions:

  1. Does the medially conforming GMK Sphere design provide an AP-stable articulation that provides for tibiofemoral translations that are comparable to, but not larger than, translations measured in natural knees?

  2. Does the medially conforming GMK Sphere design provide sufficient rotatory laxity to allow tibiofemoral rotations comparable to, but not larger than, rotations measured in natural knees?

Materials and Methods

Fifteen patients (9 females), mean age 65 years and mean BMI of 30 ±3, consented to participate. Sixteen knees received the GMK Sphere TKA. Mean Oxford Knee Score (OKS) improved significantly from 19±7 to 40±3 six months post surgery (P< 0.0001). On the day of the study, the mean OKS, Knee Society Score, EQ5D and Heath status scores were 40, 87, 0.83 and 85 respectively. Mean ROM from active maximum extension till maximum supine flexion was 108°±8°.

Motions in 16 knees were observed using pulsed-fluoroscopy during a range of activities. Subjects were observed in maximum flexion kneeling and lunging positions, and in stepping up/down on a 22cm step. Model-image registration methods were used to quantify three-dimensional knee motions from digitized fluoroscopic images.

Results

Tibial internal rotation averaged 8° during lunge and kneeling activities. During lunging, the medial and lateral condyles were an average of 2mm and 8mm posterior to the tibial sulcus, respectively, and 2mm and 9mm posterior to the tibial sulcus during kneeling. During the stair-stepping activity, the medial condyle did not translate significantly, while the lateral condyle moved 5mm posteriorly with flexion, accompanying 5° tibial internal rotation.

Discussion

The GMK Sphere TKA was designed to provide intrinsic stability through a medially conforming articulation, and provide for more natural tibial rotation with an unconstrained lateral articulation. Fluoroscopic observation of these knees during lunge, kneel and stair-stepping activities showed a stable medial articulation with little translation, and a lateral articulation translating in direct relation to tibial rotation. Tibial rotation during kneeling (8° average) was approximately twice that observed in knees with an earlier medially conforming TKA design (Moonot et al., Knee Surg Sports Traumatol Arthrosc, 2009) and similar to that observed in natural knees with medial osteoarthritis (Hamai et al., J Orthop Res, 2009). At only six months follow-up, knees with the GMK Sphere arthroplasty show functional kinematics that are AP stable and have more natural tibial rotation, consistent with the implant design intent.