Full Access

General Orthopaedics

FUNCTIONAL ACETABULAR INCLINATION ANGLE: A NEW MEASURE TO CALCULATE THE ANGLE OF THE CUP ON STANDING IN TOTAL HIP REPLACEMENTS

The Indian Orthopaedic Society (UK) (IOSUK)



Abstract

We present a study done to measure the change of angle of the acetabulum or cup, due to leg length discrepancy, deformity of hip and spine on standing. In 1998 a 3-dimensional reconstruction of hip model was prepared on CAD and the change of angle of the cup was measured as Functional Acetabular Inclination Angle (FAIA) with patient standing without squaring the pelvis. The FAIA on standing was compared with angle of the cup with patient in supine position with squared pelvis. The position of the cup changed on weight bearing due to multiple issues. The results showed that one centimetre of leg lengthening changed FAIA by 3°, 10° of abduction deformity resulted in apparent lengthening of 2.87 cm and loss of lordosis anteverted the cup on loading and vice-versa. We conclude that fixed hip deformities, leg length discrepancy and spine deformities can affect the angle of cup in hip replacement surgery and may prone to dislocations, impingement and segmental wear of the cup.