General Orthopaedics

IMPROVEMENT OF SEMI-AUTOMATED 3D KINEMATIC MEASUREMENT OF TOTAL KNEE ARTHROPLASTY FROM X-RAY FLUOROSCOPIC IMAGES

Computer Assisted Orthopaedic Surgery (CAOS) 14th Annual Meeting



Abstract

To achieve 3D kinematic analysis of total knee arthroplasty (TKA), 2D/3D registration techniques which use X-ray fluoroscopic images and computer-aided design (CAD) model of the knee implants, have been applied to clinical cases. These techniques are highly valuable for dynamic 3D kinematic measurement of TKA implants, but have needed time-consuming and labor-intensive manual operations in some process. To overcome a manual operations problem of initial pose estimation for 2D/3D registration, this study proposes an improvement method for semi-automated 3D kinematic measurement of TKA using X-ray fluoroscopic images.

To automatically estimate the initial pose of the implant CAD model, we utilise a transformation with feature points extracted from the previous and next frames. A transform matrix which has three degree of freedom (translations parallel to the image, and a rotation perpendicular to the image) is calculated by registration of corresponding feature points between the previous and next frame extracted with speeded up robust features (SURF) algorithm. While, the corresponding point sets extracted by SURF sometimes include some error sets. Therefore, in this study, least median of squares method is employed to detect the error corresponding sets and calculate a transform matrix accurately. Finally, the 3D pose of the model estimated (by the 2D/3D registration) in previous frame is transformed with the accurately calculated transform matrix, and the transformed pose is used as an initial 3D pose of the model (for the 2D/3D registration) in next frame.

To validate the feasibility of the improved semi-automated 3D kinematic measurement method, experiments using X-ray fluoroscopic images of four TKA patients during knee motions were performed. In order to assess the performance of the improved method, automation rate was calculated, and the rate was defined as the X-ray frame number of satisfying clinical required accuracy (error within 1mm, 1 degree) relative to all X-ray frame number. As results of the experiments, 3D pose of the model for all X-ray images except for the first frame is automatically stably-estimated, the automation rate of the femoral and tibial component were 83.7 % and 73.5 %, respectively.

The improved method doesn't need labor-intensive manual operations for 3D kinematic measurement of TKA, and is thought to be very helpful for actual clinical practice.