Abstract
Summary Statement
Reverse shoulder design philosophy can impact external rotation moment arms. Lateralizing the humerus can increase the external rotator moment arms relative to normal anatomy.
Introduction
The design of reverse shoulders continues to evolve. These devices are unique in that they are not meant to reproduce the healthy anatomy. The reversal of the fulcurm in these devices impacts every muscle that surrounds the joint. This study is focused on analyzing the moment arms for the rotator cuff muscles involved in internal and external rotation for a number of reverse shoulder design philosophies.
Methods
Four of the most common design philosophies were chosen. The first, a Grammont style prosthesis, with a center of rotation (COR) on the glenoid face and a humeral cup countersunk into the proximal humerus (MGMH). The second concept is the MGMH design lateralised by a 10mm bone graft (BIO). The third concept has a lateralised glenosphere COR and a humeral component inside the proximal humerus (LGMH). The fourth design has a medialised COR with a humeral component placed on top of the humerus (MGLH). This places the humerus further lateral than the previous designs. For each component set, a representative implant was modeled based on published specifications. Each design was implanted into the same digital bone models (consisting of a humerus, scapula, clavicle, and ribcage) following the manufacturer's recommended surgical technique. The muscles analyzed were the posterior-deltoid (PD), subscapularis (SSC), infraspinatus (IS), and teres minor (TM). These muscles were allowed to wrap around the bone of the scapula and proximal humerus through the range of motion. All muscle origin and insertion points were kept constant throughout the analysis. The assemblies were externally rotated from an initial position of 45° internal rotation to 45° of external rotation of the humerus with the arm at 0° of abduction. The moment arms for all muscles were compared to those calculated for the anatomic shoulder.
Results
All the rotator cuff muscles displayed a similar trend with the reverse shoulder. The external rotators all had similar moment arm values at neutral (IS∼22mm, TM∼20mm), but increased at rates proportional to their humeral offsets with external rotation (IS-MGLH 32.3mm, LGMH 27.5mm, MGMH and BIO 26.25mm; TM-MGLH 31.3mm, LGMH 27.8mm, MGMH and BIO 26.5mm). The SSC internal rotation moment arm remains roughly constant at 20mm for the anatomic shoulder, but varies widely from 45° external to 45° internal rotation with the different designs (MGLH 31.4mm to 6.7mm; MGMH 25.1mm to 11.2mm; LGMH 26.2mm to 10.8mm; BIO 25.4mm to 4.8mm). The PD moment arm is increased relative to the anatomic shoulder during external rotation for the MGLH design (9.3mm vs. 7.4mm). The other designs exhibit a decrease in the moment arm of this muscle relative to the anatomic design (LGMH 7.3mm, MGMH 5.8mm, BIO 6.4mm).
Discussion
The lateral offset between the center of humeral axis and the muscle insertion on the humerus dominates the external rotation moment arm value through this range of motion. This is evident by the increase in the moment arms with external rotation for the different reverse shoulder designs. The increase in external rotation efficiency for the external rotators and PD could play a critical role in post-operative external rotation strength and motion.