Abstract
The long-term clinical and radiological results of 63 uncemented Low Contact Stress (LCS) total knee replacements in 47 patients with rheumatoid arthritis were reviewed. The average age at the time of surgery was 69 years (53–81). At a mean follow up of 22 years (20–25), 12 patients (17 knees) were alive, 27 (36 knees) had died, and 8 patients (10 knees) were lost to follow-up.
Revision was necessary in seven patients (7 knees) (11.1%) at mean 12.1 years following surgery. Four revisions were performed due to meniscal bearing wear, two for collapse of the tibial component, and one for aseptic loosening. Evidence of post-operative infection occurred in two knees (3.2%) within 6 weeks of surgery but resolved with antibiotics. Within the group of deceased patients, five had undergone revision (included in total revisions) but otherwise the primary implant remained in vivo.
For all living patients, the mean Oxford Knee Score (/48) was 30.2 (16–41) at latest follow up at mean 19.5 years (15–24.7) following surgery. Mean active flexion was 105 degrees (90–150) at this time point.
Our recorded survival rate of the uncemented LCS total knee replacements in patients with rheumatoid arthritis was therefore 88.9% at mean 22 years, or worst-case survival of 73.0% if patients lost to follow-up were considered failures. From a review of the literature and as far as we are aware, this study represents the longest follow up of any uncemented knee arthroplasty performed in a cohort of patients with rheumatoid arthritis.