Abstract
There is continued concern over complication rates (20–30% of cases) in locked proximal humeral plating. The most common sequelae of this is screw penetration of the humeral head. This is associated with natural settling of the fracture, malreducition in varus, insufficent medial support of the fracture. The proximity of the screws to the articular surface can also be influential on outcome if collapse occurs. Our operative technique is to establish the rotation of the humeral head where the drill appears closest to the articular margin (by sequential xray screening) and subtract from this to avoid intra-articular penetration of the humeral head.
55 Consecutive patients of average age 56.4 years (14.7–86.1), 17 male and 38 females, who underwent PHILOS plating were identified using Bluespier database. Xrays were analysed for fracture pattern, restoration of neck-shaft angle, plate positioning, number and configuration of screws and presence of screw penetration both intra-operatively and at postoperative follow-up.
There were 6.07 screws used per head (total 330). There was one intraoperative screw penetration and 3 patients had evidence of screw penetration at follow-up, which required implant removal (total screw penetration rate of 7%). There was one case of AVN. The mean neck shaft angle was 137 degrees (anatomical 135 degrees).
Accurate reduction of fractures and placement of screws in the humeral head using image intensifier can act to minimise risk of screw penetration and make some of the complications of locked proximal humeral plating avoidable.