Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

Effects of Surface Treatments on Titanium-Cement Implant Bonding

International Society for Technology in Arthroplasty (ISTA) 2012 Annual Congress



Abstract

The use of polymethyl methacrylate based cement for the fixation of joint replacements although commonly applied, is still limited by interfacial weakness. This study aims to document the effects of a variety of surface treatments on implant/cement bonding and link them to their surface properties.

Thirty seven femoral implant analogues of Ti6Al4V rods were given one of six different surface treatments: traditional grit blasting, wet and dry Vaquasheening, acid etching in concentrated sulphuric and hydrochloric acid, anodisation at 150V, and a combination of acid etching and anodisation, before being embedded into a commercially available poly(methyl methacrylate) bone cement. The interfacial strength, energy and stiffness were measured through pushout testing. Surface analysis included examination with scanning electron microscopy, wettability tests and roughness analysis. Results were analysed with a one-way ANOVA with post hoc tests.

Overall, the coarse blasted surface created the strongest interface, followed by both etched then anodised, acid etched only, wet Vaquasheened, anodised only and finally dry vaquasheened. While anodised samples showed a weaker bond than etched samples, the combination of etching and anodisation was not different to etching alone. In addition, six different types of interface failure modes were observed, and theories as to explain their mechanism, using experimental evidence were outlined.

Coarse blasted surfaces showed the strongest bonding, while other surface modifications may encourage tissue ingrowth and other biological responses, these surface treatments do not strengthen bonding for cemented fixation.