Development of a Novel Knee Simulator to Evaluate Kinematics and Kinetics of TKA

International Society for Technology in Arthroplasty (ISTA) 2012 Annual Congress



Abstract

We have developed a novel knee simulator that reproduces the active knee motion to evaluate kinematics and joint reaction forces of TKA.

There have been developed many kinds of knee simulators; Most of them are to predict TKA component wear and the others are to evaluate the kinematics and/or kinetics of TKA. The most simulators have been operated using the data of the loading and kinematics profile of the knee obtained from normal gait. Here a problem is that such variables as joint force and kinematics are the outcome caused by the application of muscles' and external forces. If so, a simulator should be operated by the muscles' and external forces so as to duplicate the in vivo condition. Other disadvantages for the current knee simulators are; a knee joint motion is made passively, the effects of the hip joint motion are not taken into account, and the maximum flexion angle is usually limited at about 100°.

Considering the above, we have developed a knee simulator with the following advantages and innovative features. First, the simulator is driven by the muscles' forces and an active knee motion is made with bearing the upper body weight. As a result, the knee shows a 3D kinematics and generates the tibio-femoral contact forces. Under this condition, the TKA performance is to be assessed. Secondly, a hip joint mechanism is also incorporated into the simulator. The lower limb motion is achieved by the synergistic function between the hip and knee joints. Under this condition, a natural knee motion is to be reproduced. Thirdly, the simulator can make complete deep knee flexion up to 180°. Thus not only the conventional TKA but also a new TKA for high flexion can be attached to it for the evaluation.

Figure 1 shows the structure of the simulator, in which both the hip and knee joints are moved in a synergistic fashion by the pull forces of four wires. The four wires are pulled by the four servomotors respectively and reproduce the functions of the mono-articular muscles ((1), (3)) and the bi-articular muscles ((2), (4)) through the multiple pulley system. It should be noted that weight A and B are not heavy enough for the inverted double pendulum to stand up straight. They are applied as counter weights so that each segment duplicate the each segmental weight of the human lower limb. Figure 2 shows a sequential representation of stand to sit features: (a) at standing, (b) at high flexion, and (c) at deep flexion. At a state of 130° knee flexion between (b) and (c), hamstrings wire (4) becomes shortest and then exhibits an eccentric contraction, thereby attaining deep flexion.

Our knee simulator can be a useful tool for the evaluation of TKA performance and may potentially substitute the in vivo experiments.