Abstract
Implant alignment in knee arthroplasty has been identified as critical factor for a successful outcome. Human error during the registration process for imageless computer navigation knee arthroplasty directly affects component alignment. This cadaveric study aims to define the error in the registration of the landmarks and the resulting error in component alignment.
Five fresh frozen cadaveric limbs including the hemipelvis were used for the study. Five surgeons performed the registration process via a medial parapatellar approach five times. In order to identify the gold standard point, the soft tissues were stripped and the registration was repeated by the senior author. Errors are presented as mm or degrees from the gold standard registration.
The error range in the registration of the femoral centre in the coronal plane was 6.5mm laterally to 5.0mm medially (mean: −0.1, SD: 2.7). This resulted in a mechanical axis error of 5.2 degrees valgus to 2.9 degrees varus (mean: 0.1, SD: 1.1). In the sagittal plane this error was between −1.8 degrees (extension) and 2.7 degrees (flexion). The error in the calculation of the tibial mechanical axis ranged from −1.0 (valgus) to 2.3 (varus) degrees in the coronal plane and −3.2 degrees of extension to 1.3 degrees of flexion. Finally the error in calculating the transepicondylar axis was −11.2 to 6.3 degrees of internal rotation (mean: −3.2, SD: 3.9).
The error in the registration process of the anatomical landmarks can result in significant malalignment of the components. The error range for the mechanical axis of the femur alone can exceed the 3 degree margin that has been previously been associated with implant longevity. The technique during the registration process is of paramount importance for image free computer navigation. Future research should be directed towards simplifying this process and minimizing the effect of human error.