Long term follow-up of the agility total ankle replacement

British Orthopaedic Association 2012 Annual Congress



Abstract

We aimed to review the outcome of Agility total ankle replacements carried out in our institution between 2002 and 2006.

Follow-up consisted of clinical and radiological review pre-operatively, then at 6 weeks, 6 and 12 months, and annually until 10 years post op. Clinical review included the American Orthopaedic Foot and Ankle Score, satisfaction and pain scores. Case notes were reviewed to determine intra and post-operative complications.

30 arthroplasties were performed in 30 consecutive patients. Pre-operative diagnosis was rheumatoid arthritis(16), primary osteoarthritis(12) and post-traumatic osteoarthritis(2). After a mean follow up of 6.2 years (1.4–10.1), 4 patients had died, and 22 out of the remaining 24 were available for follow-up. Intra operative complications included lateral malleoli fracture(3) and superficial peroneal nerve injury(2). Post operative complications included 1 early death, but this was not related to the surgical procedure. Two patients developed deep infections of the prosthesis. One underwent removal of the implant; the other is on long term oral antibiotic therapy. One patient had delayed union of the syndesmosis and six patients had non-union. On clinical assessment, patients' AOFAS scores improved from mean 40.4 pre-op to 83.5 post-op (p< 0.001). Radiological assessment of the tibial component revealed 25 (93%) patients had lucency in at least one zone in the AP radiograph.

We found a relatively high level of re-surgery and complications following Agility total ankle replacement. A 7% revision rate is much higher than would be tolerated in knee or hip arthroplasty, but compares favourably to other studies of TAR. Despite radiological features which suggest loosening, the high rate of re-surgery and complications; patients are generally satisfied with the procedure, reporting lower levels of pain and improved function. Overall we feel that the Agility ankle is an acceptable alternative to ankle arthrodesis, however patients need to be warned of the risk of re-surgery.