Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

CONTRALATERAL ELBOW RADIOGRAPHS CAN RELIABLY DIAGNOSE RADIAL HEAD IMPLANT OVER-LENGTHENING

Canadian Orthopaedic Association (COA)



Abstract

Purpose

Radial head implant over-lengthening, a common cause of capitellar wear and clinical failure, is difficult to diagnose using radiographs of the injured elbow. The purpose of this study was to determine if a novel measurement technique based on contralateral elbow radiographs, termed the RACER method, could be used to accurately estimate the magnitude of radial head implant over-lengthening. Part I of this study examined the side-to-side consistency of radiographic landmarks used in the measurement technique. Part II of this study validated the technique using simulated radial head implant over-lengthening in a cadaveric model.

Method

Part I: A side-to-side comparison of elbow joint dimensions was performed in 50 patients (100 radiographs). Part II: Radial head prostheses of varying lengths (0,+2mm,+4mm,+6mm,+8mm) were implanted in 4 paired cadaveric specimens (8 elbows). Radiographs were obtained and measurements were performed by 2 examiners blinded to implant size to determine if contralateral radiographs could diagnose and provide a valid estimate of the magnitude of implant over-lengthening. Intra and inter-rater reliability was determined.

Results

No significant side-to-side differences (p>0.2) in radiographic measurements were identified between paired elbows. The RACER measurement technique using contralateral radiographs was successful in predicting the implant size (1mm) in 104 of 120 (87%) of scenarios tested. The sensitivity of the technique, the ability of the test to correctly identify over-lengthening when it was present within 1mm, was 98%. The intra-rater agreement for a single orthopaedic surgeon measuring on 2 separate occasions was excellent (ICC>0.90). The inter-rater agreement between two separate surgeons was also excellent, with the 95% lower confidence interval exceeding 0.90 in all cases.

Conclusion

A novel measurement technique based on contralateral elbow radiographs can be used to diagnose and calculate the magnitude of radial head implant over-lengthening. Clinical Significance: Implantation of an incorrectly sized radial head prosthesis is not uncommon. The described technique can be used to effectively diagnose and determine the magnitude of over-lengthening.