Abstract
Limb lengthening with external fixators has many complications, like pin track infection, joint stiffness, severe pain and stress fracture after removal of external fixator. Prolonged period in external fixator interferes with rehabilitation and activities of daily living.
We describe our results with ISKD (intramedullary skeletal kinetic distractor) system, an internal limb lengthening device, activated with rotations of 3 degrees. In this retrospective cohort we performed lengthening with the use of ISKD in 12 patients from March 2006 to date. Ten patients were included in this study as two patients had not completed the stage of consolidation. There were 7 male and 3 female patients. Their average age was 44 years (range 23–63). The mean follow up period was 16 months (range 12–24 months). Two patients had tibial lengthening and eight patients had femoral lengthening. The average lengthening desired was 42.5 mm (range 2.5 mm–75 mm). The average lengthening achieved was 38 mm. Mean hospital stay was 7 days (range 5–11 days). The distraction index was 1.2 mm/day. The average time to full weight bearing was 6 months (range 4 to 10 months). Mean healing time was 10.25 months. The mean healing index was 90.7 days/cm. If we remove two patients who took long time to heal, the mean healing index drops to 61 days/cm. Two patients needed bone grafting at the distraction site. None of the patients had infection, non-union of the distraction site or breakage of the nail. None of the patients had joint stiffness. All patients were completely satisfied with the treatment and had excellent functional results. We conclude that the low rate of complications, higher patient acceptability and avoidance of external fixation make ISKD a very attractive option for limb lengthening.