Large diameter metal-on-metal articulation total hip replacement versus 28mm metal-on-polyethylene - is there a functional difference at 5 yrs?

British Orthopaedic Association/Irish Orthopaedic Association Annual Congress (BOA/IOA)



Abstract

Introduction

Large head total hip arthroplasty (THA) reduces dislocation rates and provides a theooretically larger range of motion. We hypothesised that this would translate into greater improvement in functional scores when compared to 28mm metal-on-polyethylene THA at 5 years. We believe ours to be the first in vivo comparison study.

Methods

A multi-surgeon case-control study in a District General Hospital. The study group consisted of 427 patients with 452 hips, the 38mm uncemented metal-on-metal articulation THA (M2A/Bi-metric, Biomet UK). The control group consisted of 438 age and sex-matched patients with 460 28mm metal-on-polyethylene articulation THA (Exeter/Exeter or Exeter/Duraloc - Stryker UK. All patients were assessed in a physiotherapist led Joint Review Service as part of their standard follow up, with functional scoring using Oxford Hip (scored 0–48) and WOMAC scores (0–100).

Results

The demographics for the 38mm and 28mm groups gave mean ages of 65.8 years and 66.4 years, 40.4% and 39.3% male respectively. Pre-operative functional scores were comparable, with Oxford Hip scores of 23.3 and 26.8 respectively, WOMAC 49 compared to 53. At each review point there was no statistical difference in either Oxford or WOMAC scores and this was sustained at 5 yrs. Dislocation rates in the 38mm group were lower (2.9% vs. 5%) though not statistically significant (p = 0.111). Revision rate was significantly higher in the larger head group, primarily due to adverse reaction to metal debris (4.6% vs. 2.0%)

Conclusions

There is no functional difference between 38mm metal-on-metal THA and 28mm metal-on-polyethylene THA at five years. Dislocation rates were found to be lower in the 38mm THAs as would be expected, but this was not statistically significant. The difference in revision rates was found to be due to metal-on-metal adverse reactions to metal debris, and their use is therefore not advocated in the current climate