Abstract
Femoral component fracture is a rarely reported but devastating complication of total knee arthroplasty. It has occurred most frequently with Whiteside Ortholoc II replacements uncemented knee replacements. Presentation may be with acute pain, progressive pain or returning deformity. It occurs more commonly in the medial condyle of the femoral component. It is rarely seen in cemented replacements.
All currently available literature describing fractures of condylar replacements, both cemented and uncemented. Predisposing factors include varus deformity either pre or post operatively. The mechanism of failure is thought to be failure of the infiltration of bone into the replacement. This is often due to polyethylene wear or metallosis causing abnormal tissue reaction with or without osteolysis.
We present the case of a fractured Press Fit Condylar (PFC) cemented implant (DePuy, Johnson&Johnson, Raynham, Massachusettes, USA) affecting the medial condyle. To our knowledge this is only the third reported case of fracture in a PFC implant, and the first in a cemented PFC implant.
Our patient was a 64 year old male who presented with unresolving knee pain post total knee arthroplasty, caused by fatigue fracture of the medial condyle of the femoral component. This was identified as loosening on plain radiographs and replaced with a revision prosthesis with a good post operative result. Given our aging population and with the increase of joint arthroplasty, this case sheds light on a potentially under recognised and increasingly important cause of knee pain following arthroplasty.