Abstract
The aim of this retrospective study was to review the outcome of patients treated with Fassier-Duval (FD) rods and highlight some of the complications found during treatment.
Between April 2006 and August 2010 we inserted 24 FD rods in 13 patients. 17 rods for osteogenesis imperfecta (OI), 2 for fractures and deformity associated with cerebral palsy, 1 for fracture associated with muscular dystrophy, 1 for fibrous dysplasia and 3 for centralisation of single bone forearms.
In the upper limb one patient required revision for proximal migration of the male component and another patient is waiting for revision for the same problem.
In the lower limb, a tibial nail was revised because of proximal migration of the male component. A femoral nail was adjusted because of loss of the proximal fixation. One of the OI patients fell, fractured the femur and bent a femoral nail. This awaits revision at a later date. A second OI patient fell on 2 separate occasions bending both a tibial and a femoral nail respectively. These were both revised to trigen intramedullary nails.
In all the other cases there were no complications.
In summary the Fassier Duval system provides a versatile way of providing intramedullary stabilisation for growing bones through a single entry point. However in our experience we have a 33% complication rate most notably bending of the rods. We advocate careful patient selection and using as high a diameter nail as is feasible.