LONG-TERM RESULTS OF CEMENTLESS FULLY HAC COATED ACETABULAR CUPS IN PRIMARY HIP ARTHROPLASTY

Australian Orthopaedic Association Limited (AOA)



Abstract

We report the clinical and radiological outcome of consecutive primary hip arthroplasties using the JRI-Furlong Hydroxyapatite ceramic (HAC) coated acetabular components.

We reviewed 412 consecutive cementless primary THA using fully coated acetabular shell in 392 patients—with a minimum 12 to 18 year follow-up—performed at two institutions between 1986 and 1994. Twenty (22 THA) were lost prior to 12-year follow-up, leaving 372 patients (390 THA) available for study. Fully HAC coated stems were used in all patients. The clinical outcome was measured using Harris, Charnley and Oxford hip scores and the quality of life using EuroQol EQ-5D. Radiographs were systematically analysed for implant position, loosening, migration, osteolysis. Polythene wear was digitally measured. The radiographic stability of the acetabular component was determined by Enghs criteria.

The mean age was 74.4 years. The mean Harris and Oxford scores were 87 (78– 97) and 19.1 (12–33) respectively. The Charnley score was 5.6 (5-6) for pain, 5.2 (4–6) for movement and 5.3 (4–6) for mobility. Migration of acetabular component was seen in four hips. Acetabular radiolucencies were present in 54 hips (9.7%). The mean linear polythene wear was 0.06mm/year. Mean inclination was 48.4° (38–65).

Radiolucencies were present around 37 (6.6%) stems. Dislocation occurred in 10 patients (three recurrent). Re-operations were performed in nine patients (1.9%). Four acetabular revisions were performed for aseptic loosening. Other re-operations were for infection (three), periprosthetic fractures (one), cup malposition (one) and revision of worn liner (three). Mean EQ-5D description scores and health thermometer scores were 0.81 (0.71–0.89) and 86 (64–95). With an end point of definite or probable loosening, the probability of survival at 12 years was 97.1% for acetabular component. Overall survival at 12 years with removal or repeat revision of either component for any reason as the end point was 96.2%.

The results of this study support the continued use of a fully coated prosthesis and documents the durability of the HAC coated components. In our clinical experience, the Furlong prosthesis revealed encouraging radiographic stability over a long-term period.