FINITE ELEMENT MODELLING OF THE BIRMINGHAM HIP MID HEAD RESECTION

Australian Orthopaedic Association Limited (AOA)



Abstract

The Birmingham Hip Mid Head Resection (BMHR) was designed to accommodate patients with lower quality bone in the proximal half of the femoral head. It is a relatively new conservative hip implant with promising early results. Finite element modelling may provide an insight into mid-term results.

A cadaveric femur was CT scanned and 3D geometry of the intact femur constructed. The correctly sized BMHR implants (with and without visual stop) were positioned and these verified by a surgeon; hence constructing the post-operative models. Walking loads were applied and contact surfaces defined.

Stress analyses were performed using the finite element method and contact examined. Also, a strain-adaptive bone remodelling analysis was run using 45% gait hip loading data. Virtual DEXA images were computed and were analysed in seven regions of the bone surrounding the implants.

The BMHR was found to be mechanically stable with all surfaces indicating micromotion less than the critical 150 microns. Stress distribution was similar to the intact femur, with the exception of the head-neck region where some stress/strain shielding occurs. This is mirrored in the bone remodelling results, which show some bone resorption in this region. The visual stop, which is designed to ensure that the stem is not overdriven during implantation, did not affect the stress/strain results; only on a very local scale.

There is minimal data available in the literature regarding conservative hip implants and no data regarding the BMHR. This study is the first to look at the mechanical response of the bone to this implant.