PRECLINICAL WORST-CASE TESTING OF A NOVEL VITAMIN-DOPED HXLPE

Australian Orthopaedic Association and New Zealand Orthopaedic Association (AOA/NZOA) - Combined Annual Scientific Meeting



Abstract

The introduction of a new implant material is not without risk. A series of worst-case scenarios were developed and tested accordingly to answer questions such as: what will happen if the implant is not placed in a good orientation? What will happen to the material after a long implantation time, e.g. 20 or more years?

To reach a higher level of safety, a new approach for the preclinical testing has been taken. The vitamys® material (a novel vitamin-doped HXLPE) followed a severe pre-clinical testing protocol, including mechanical, tribological and biocompatibility testing. The testing includes a comparison of vitamys® vs. standard-UHMWPE and other HXLPE after accelerated ageing for periods equivalent to 20 and 40 years in-vivo. Hip simulator testing was done at inclination angles from 35° to 65° to assess the “forgiveness” of the material for mal-orientation.

Comparing the test results to published data, it becomes evident that the vitamin addition and the sequence of the manufacturing steps both have a significant effect of the resulting mechanical, ageing and wear properties. In contrast to UHMWPE or HXLPE without antioxidant, the vitamys material behaves in a very “forgiving” manner: Hip simulator testing of vitamys at high inclination angles and even with severely aged material revealed no increase of wear rates. The vitamys material was first introduced in a monoblock polyethylene cup with a thin Ti-particle coating, the RM-Pressfit vitamys® acetabular cup (Mathys Ltd Bettlach, Switzerland). Its first implantation occurred in Sept. 2009. Since then, a total of nearly 500 implantations have been documented in a prospective multi-centre clinical study involving 11 clinics in 5 countries (CH, DE, FR, NL and NZ).

Based on the pre-clinical testing and its first clinical experience, we have reason to believe that the RM-Pressfit vitamys® possesses interesting and unique features such as high elasticity (no stress-shielding), high ageing and wear resistance combined with clinically proven biological anchorage – making it theoretically suitable for a whole range of patients, including the young and active.