VALIDATION OF A NEW TENSION SYSTEM FOR ACL RECONSTRUCTION

British Orthopaedic Association (BOA) 2005



Abstract

The aim of this paper is to evaluate the Linvatec SE graft tensioner system in obtaining predictable initial tension during ACL reconstruction using hamstring grafts.

The gracilis and semitendinosus grafts were tensioned individually prior to fixation distally to a combined tension of 80N. The knees were then cycled through full range of motion and the tension recorded at 90 degrees and in full flexion and extension. Experience on the first 22 patients indicated that 41% required 10 cycles of the knee to remove pre-conditioning and equalise tension at 80N combined, while the remaining 50% required 15 cycles. 90% reached stable tension after 15 cycles. In 54% the tension increased at full extension and graft tension was adjusted to 80N in extension to avoid overconstraining the graft. Twenty-two patients studied following this initial protocol underwent outcome assessment after minimum 6 months. Mean KT1000 arthrometry manual maximum side to side difference was 1.5mm. Femoral fixation was achieved using the Endobutton (Smith and Nephew) and Tibial fixation using the Extralok bioabsorbable screw (Linvatec).

A subsequent shortened protocol of initial over-tensioning to 60N and 40N for the semitendinosus and gracilis double bundles respectively, followed by 15 cycles of the knee resulted in stable 80N combined tension with no further drop with more cycling.

We conclude that the new tensioner system allows for accurate and predictable initial tension of hamstring reconstructed ACL grafts and that its use can be simplified by using the second technique protocol.