Research

THE PREOPERATIVE GAIT PATTERN IS ASSOCIATED WITH MIGRATION OF TOTAL KNEE ARTHROPLASTY: AN EXPLORATORY RADIOSTEREOMETRY STUDY WITH TWO YEARS’ FOLLOW-UP

The European Orthopaedic Research Society (EORS) 32nd Annual Meeting, Aalborg, Denmark, 18–20 September 2024.



Abstract

Introduction

Knee osteoarthritis often causes malalignment and altering bone load. This malalignment is corrected during total knee arthroplasty surgery, balancing the ligaments. Nonetheless, preoperative gait patterns may influence postoperative prosthesis load and bone support. Thus, the purpose is to investigate the impact of preoperative gait patterns on postoperative femoral and tibial component migration in total knee arthroplasty.

Method

In a prospective cohort study, 66 patients with primary knee osteoarthritis undergoing cemented Persona total knee arthroplasty were assessed. Preoperative knee kinematics was analyzed through dynamic radiostereometry and motion capture, categorizing patients into four homogeneous gait patterns. The four subgroups were labeled as the flexion group (n=20), the abduction (valgus) group (n=17), the anterior drawer group (n=10), and the tibial external rotation group (n=19). The femoral and tibial component migration was measured using static radiostereometry taken supine on the postoperative day (baseline) and 3-, 12-, and 24- months after surgery. Migration was evaluated as maximum total point motion.

Result

Of the preoperatively defined four subgroups, the abduction group with a valgus-characterized gait pattern exhibited the highest migration for both the femoral (1.64 mm (CI95% 1.25; 2.03)) and tibial (1.21 mm (CI95% 0.89; 1.53)) components at 24-month follow-up. For the femoral components, the abduction group migrated 0.6 mm (CI95% 0.08; 1.12) more than the external rotation group at 24 months. For the tibial components, the abduction group migrated 0.43 mm (CI95% 0.16; 0.70) more than the external rotation group at 3 months. Furthermore, at 12- and 24-months follow-up the abduction group migrated 0.39 mm (95%CI 0.04; 0.73) and 0.45 mm (95%CI 0.01; 0.89) more than the flexion group, respectively.

Conclusion

A preoperative valgus-characterized gait pattern seems to increase femoral and tibial component migration until 2 years of follow-up. This suggests that the implant fixation depends on load distributions originating from specific preoperative gait patterns.


Corresponding author: Emil Toft Petersen