header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Foot & Ankle

WHAT HAPPENS AFTER FAILED TOTAL ANKLE ARTHROPLASTY?

The British Orthopaedic Foot & Ankle Society (BOFAS) Annual Congress 2024, Belfast, Northern Ireland, 6–8 March 2024.



Abstract

Aims

Retrospective review of a consecutive series of 1,168 total ankle replacements (TAR) performed at Wrightington, to analyse modes of failure and clinical outcomes following TAR failure.

Methods

All patients undergoing TAR between November 1993 – June 2019 were collated (4–25 year follow-up; mean 13.7 years). 6 implants were used (300 STAR, 100 Buechal Pappas, 509 Mobility, 118 Zenith, 41 Salto and 100 Infinity). 5 surgeons, all trained in TAR, performed the surgery. Modes of failure were collated and clinical and radiological outcomes recorded for the revisional surgery following failure of the TAR.

Results

156 (13.4%) TARs failed (47STAR 15.6%, 16BP 16%, 77Mobility 15.1%, 6Salto 14.6%, 10Zenith 8.5% and 0Infinity 0%). Mean time to failure 5.8 years (0.1- 21.4 years). The 4 most common modes of failure were 44.9% aseptic loosening, 11.5% gutter pain, 10.9% infection and 10.3% recurrent edge loading. 50 underwent conversion to tibiotalocalcaneal (TTC) fusion with nail with 9 (18%) failing to fuse. 31 underwent revision TAR with 2 (6.5%) subsequently failed. 22 underwent ankle fusion with 10 (45%) failing to fuse. 21 underwent polyethylene exchange of which 8 (38%) had further poly failure. 20 (12.8%) were managed conservatively, 2 (1.3%) required below knee amputation and 6 were listed but lost to follow-up. 81 of the 1168 (7%) consecutive cohort were lost to follow-up.

Conclusions

13.4% of the TAR cohort have failed at average follow-up 13.7 years. There was no difference in failure modes across the implant designs. Whilst the fixed bearing has the shortest follow-up, it may be performing better as there have been no failures so far. Prior to October 2016, most revisions were to fusion (TTC 18% failure rate, ankle 45% failure rate), whereas post 2016, 57% patients elected for revision TAR (6.5% failure).