DEVELOPMENT AND VALIDATION OF A CLINICAL TOOL TO SEMI-AUTOMATIC MEASURE TAR ALIGNMENT AND MIGRATION ON RADIOGRAPHS

The International Combined Orthopaedic Research Societies (ICORS), World Congress of Orthopaedic Research, Edinburgh, Scotland, 7–9 September 2022. Part 2 of 3.



Abstract

Malalignment is often postulated as the main reason for the high failure rate of total ankle replacements (TARs). Only a few studies have been performed to correlate radiographic TAR malalignment to the clinical outcome, but no consistent trends between TAR alignment parameters and the clinical outcome were found. No standard TAR alignment measurement method is present, so reliable comparison between studies is difficult. Standardizing TAR alignment measurements and increasing measurable parameters on radiographs in the clinic might lead to a better insight into the correlation between malalignment and the clinical outcome. This study aims to develop and validate a tool to semi-automatic measure TAR alignment, and to improve alignment measurement on radiographs in the clinic.

A tool to semi-automatically measure TAR alignment on anteroposterior and lateral radiographs was developed and used by two observers to measure TAR alignment parameters of ten patients. The Intraclass Coefficient (ICC) was calculated and accuracy was compared to the manual measurement method commonly used in the clinic.

The tool showed an accuracy of 76% compared to 71% for the method used during follow-up in the clinic. ICC values were 0.94 (p<0.01) and higher for both inter-and intra-observer reliability.

The tool presents an accurate, consistent, and reliable method to measure TAR alignment parameters. Three-dimensional alignment parameters are obtained from two-dimensional radiographs, and as the tool can be applied to any TAR design, it offers a valuable addition in the clinic and for research purposes.