Research

ROBOTIC ARM-ASSISTED SYSTEM FOR UNICOMPARTMENTAL KNEE ARTHROPLASTY IS ASSOCIATED WITH A LEARNING CURVE OF 11 CASES AND INCREASED OPERATING TIME

The International Combined Orthopaedic Research Societies (ICORS), World Congress of Orthopaedic Research, Edinburgh, Scotland, 7–9 September 2022. Part 2 of 3.



Abstract

Unicompartmental knee arthroplasty (UKA) has a higher risk of revision than total knee arthroplasty, particularly for low volume surgeons. The recent introduction of robotic-arm assisted systems has allowed for increased accuracy, however new systems typically have learning curves. The objective of this study was to determine the learning curve of a robotic-arm assisted system for UKA.

Methods A total of 152 consecutive robotic-arm assisted primary medial UKA were performed by five surgeons between 2017 and 2021. Operative times, implant positioning, reoperations and patient-reported outcome measures (PROMS; Oxford Knee Score, EuroQol-5D, and Forgotten Joint Score) were recorded.

There was a learning curve of 11 cases with the system that was associated with increased operative time (13 minutes, p<0.01) and improved insert sizing over time (p=0.03). There was no difference in implant survival (98.2%) between learning and proficiency phases (p = 0.15), and no difference in survivorship between ‘high’ and ‘low’ usage surgeons (p = 0.23) at 36 months. There were no differences in PROMS related to the learning curve. This suggested that the learning curve did not lead to early adverse effects in this patient cohort.

The introduction of a robotic-arm assisted UKA system led to learning curves for operative time and implant sizing, but there was no effect on patient outcomes at early follow- up. The short learning curve was independent of UKA usage and indicated that robotic-arm assisted UKA may be particularly useful for low-usage surgeons.