header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

HA-COATED SCREW VERSUS TITANIUM TRIANGULAR DOWEL FOR SACROILIAC JOINT FUSION: RESULTS FROM A PROSPECTIVE OUTCOME STUDY

The Welsh Orthopaedic Society (WOS) Meeting, Newport, Wales, 26–27 May 2022.



Abstract

Background

Prospective study to compare patient reported outcome measures (PROM) for sacroiliac joint (SIJ) fusion using HA-coated screw (HACS) vs triangular titanium dowel (TTD). First study of its kind in English literature.

Methods

40 patients underwent percutaneous SIJ stabilisation using HACS and TTD was used in 70 patients. Patients were followed up closely and outcome scores were collected prospectively. PROMs were collected preoperatively and 12 months after surgery. Short Form (SF)-36, Oswestry Disability Index, EuroQol-5D-5L and Majeed Pelvic Scores were collected. Shapiro-wilk test was used to determine normality of data. Mann-whitney U test was used to compared non-parametric data and Independent sample T test for parametric data.

Results

33 patients in HACS group and 61 in TTD group completed follow-up. There was no significant difference in all preoperative PROMs in both groups hence the data was considered comparable. All postoperative PROMs were significantly higher in the TTD group. In the HACS group, 21 patients (63%) had lysis around the screw and a sub-group analysis showed that improvement in PROMs was significantly less in patients with lysis around the screw. 4 patients with lysis around the screw were offered revision due to ongoing pain. Revision was successful in only 1 patient. In TTD group only 5 patients (8.2 %) patients had radiological evidence of lysis. Relative risk of developing lysis was 6.7 times higher in HACS group.

Conclusion

Percutaneous SIJ fixation procedure has been shown to have good clinical outcomes and TTD leads to significantly better patient reported outcomes compared to HACS. There is a 6.7 time higher risk of lysis with use of HACS and lysis is a risk factor for poor outcomes.