Abstract
Total Knee Arthroplasty (TKA) is an effective surgery performed for knee osteoarthritis. Despite this success up to 20% of patients are less than fully satisfied. Rotating platform (RP) bearings have demonstrated significant reduction in rates of wear in laboratory studies but thus far have failed to demonstrate a clinical difference compared to fixed bearing (FB) arthroplasty in patients. This may be due to studies limited by small sample size and single centre design. Additionally, no studies have utilised modern patient-reported outcome measures (PROMs) such as PKIP to investigate differences between these bearings.
A non-randomised, prospective, multi-centre study was conducted across centres in Australia, United Kingdom and the United States from November 2012 to May 2015. Knee prosthesis included were the RP and FB models of the PFC Sigma Knee (Depuy, Synthes) and the ATTUNE Knee (Depuy, Synthes). Data was collected on pre-arthroplasty interventions, demographics and PROMs. PROMs included were the KOOS, OKS, EQ53DL and PKIP. Participants were followed-up for 2 years.
A total of 1,718 patients were recruited across 27 centres. 959 participants underwent a FB TKA, 759 participants underwent a RP TKA. No significant difference was found between RP and FB in change from baseline at 2 years across any parameter in all PROMs investigated at 2 years. Body Mass Index pre-operatively was the biggest indicator of change from baseline at 2 years.
Our data demonstrates similar improvements in the short-term in both RP and TKA across all PROMs. Patients reported similar scores in satisfaction, confidence, mobility and stability in the PKIP PROM for both RP and FB groups. We have not found any evidence to suggest that RP TKA is disadvantageous at 2 years, and this is reassuring for any surgeon looking to utilise this bearing type in practice.
Our data does not demonstrate a clinical advantage in RP TKA over FB TKA at 2 years in any PROM.