Full Access

General Orthopaedics

EARLY REVISION RATES OF TOTAL HIP ARTHROPLASTY USING INTELLIJOINT HIP® COMPUTER NAVIGATION SYSTEM: A STUDY FROM THE AUSTRALIAN NATIONAL JOINT REPLACEMENT REGISTRY OF 1,911 PROCEDURES

The New Zealand Orthopaedic Association and the Australian Orthopaedic Association (NZOA AOA) Combined Annual Scientific Meeting, Christchurch, New Zealand, 31 October – 3 November 2022. Part 1 of 2.



Abstract

Total hip arthroplasty (THA) is an effective treatment for symptomatic hip osteoarthritis (OA). Computer-navigation technologies in total knee arthroplasty show evidence-supported survivorship advantages and are used widely. The aim of this study was to determine the revision outcome of hip commercially available navigation technologies.

Data from the Australian Orthopaedic Association National Joint Replacement Registry from January 2016 to December 2020 included all primary THA procedures performed for osteoarthritis (OA). Procedures using the Intellijoint HIP® navigation were identified and compared to procedures inserted using ‘other’ computer navigation systems and to all non-navigated procedures. The cumulative percent revision (CPR) was compared between the three groups using Kaplan-Meier estimates of survivorship and hazard ratios (HR) from Cox proportional hazards models, adjusted for age and gender. A prosthesis specific analysis was also performed.

There were 1911 procedures that used the Intellijoint® system, 4081 used ‘other’ computer navigation, and 160,661 were non-navigated. The all-cause 2-year CPR rate for the Intellijoint HIP® system was 1.8% (95% CI 1.2, 2.6), compared to 2.2% (95% CI 1.8, 2.8) for other navigated and 2.2% (95% CI 2.1, 2.3) for non-navigated cases. A prosthesis specific analysis identified the Paragon/Acetabular Shell THAs combined with the Intellijoint HIP® system as having a higher (3.4%) rate of revision than non-navigated THAs (HR = 2.00 (1.01, 4.00), p=0.048).

When this outlier combination was excluded, the Intellijoint® system group demonstrated a two-year CPR of 1.3%. There was no statistical difference in the CPR between the three groups before or after excluding Paragon/Acetabular Shell system.

The preliminary data presented demonstrate no statistical difference in all cause revision rates when comparing the Intellijoint HIP® THA navigation system with ‘other’ navigation systems and ‘non-navigated’ approaches for primary THAs performed for OA. The current sample size remains too small to permit meaningful subgroup statistical comparisons.