ASSESSMENT OF HEMIARTHROPLASTY IN AN ANATOMICAL HIP SIMULATOR

The British Orthopaedic Research Society (BORS) 2023 Meeting, Cambridge, England, 25–26 September 2023.



Abstract

Abstract

BACKGROUND

Hemi-arthroplasty (HA) as a treatment for fractured neck of femur has slightly increased since 2019 and remarkably after the COVID pandemic. The main drawback of the treatment is ongoing cartilage deterioration that may require revision to THR.

OBJECTIVE

This study assessed cartilage surface damage in hip HA by reproducing anatomical motion and loading conditions in a hip simulator.

METHODS

Experimental design. HA tests were conducted using porcine acetabula and CoCr femoral heads. Five groups (n=4) were included: a control group comprising natural tissue and four HA groups where the acetabula were paired with metal heads to allow radial clearance (RC) classed as small (RC<0.6mm), large (2mm<RC<4mm), extra-large (4mm<RC), and oversized (RC<−0.6mm). Tests were carried out in an anatomical hip simulator that reproduced a simplified twin peak gait cycle, adapted for porcine hip joints, from the ISO 14242 standard for wear of THR prostheses (peak load of 900N). The test length was 6 hours, with photogrammetry taken at 1-hour intervals. Ringers solution was used as a lubricant.

RESULTS

No changes were observed in the control group. However, cartilage surface changes were observed in all hemi-arthroplasty groups. Discolouration on the cartilage surface was noticeable at the posterior-superior part of the acetabulum after 1-hour (extra-large and oversized groups). Damage severity and location were characteristic of each clearance group. Of all the groups, the oversized group showed more significant damage. No labrum separation was seen after the simulation.

CONCLUSIONS

These results are relevant to understand the effect of femoral head clearance on cartilage damage risk after HA.