ACCURACY OF MEASURING IN VIVO SYNDESMOSIS JOINT KINEMATICS DURING LEVEL GAIT

The British Orthopaedic Research Society (BORS) 2023 Meeting, Cambridge, England, 25–26 September 2023.



Abstract

Abstract

Objectives

The syndesmosis joint, located between the tibia and fibula, is critical to maintaining the stability and function of the ankle joint. Damage to the ligaments that support this joint can lead to ankle instability, chronic pain, and a range of other debilitating conditions. Understanding the kinematics of a healthy joint is critical to better quantify the effects of instability and pathology. However, measuring this movement is challenging due to the anatomical structure of the syndesmosis joint. Biplane Video Xray (BVX) combined with Magnetic Resonance Imaging (MRI) allows direct measurement of the bones but the accuracy of this technique is unknown. The primary objective is to quantify this accuracy for measuring tibia and fibula bone poses by comparing with a gold standard implanted bead method.

Methods

Written informed consent was given by one participant who had five tantalum beads implanted into their distal tibia and three into their distal fibula from a previous study. Three-dimensional (3D) models of the tibia and fibula were segmented (Simpleware Scan IP, Synopsis) from an MRI scan (Magnetom 3T Prisma, Siemens). The beads were segmented from a previous CT and co-registered with the MRI bone models to calculate their positions. BVX (125 FPS, 1.25ms pulse width) was recorded whilst the participant performed level gait across a raised platform. The beads were tracked, and the bone position of the tibia and fibula were calculated at each frame (DSX Suite, C-Motion Inc.). The beads were digitally removed from the X-rays (MATLAB, MathWorks) allowing for blinded image-registration of the MRI models to the radiographs. The mean difference and standard deviation (STD) between bead-generated and image-registered bone poses were calculated for all degrees of freedom (DOF) for both bones.

Results

The absolute mean tibia and fibula bone position differences (Table 1) between the bead and BVX poses were found to be less than 0.5 mm for both bones. The bone rotation differences were found to be less than 1° for all axes except for the fibula Z axis rotation which was found to be 1.46°. One study1 has reported the kinematics of the syndesmosis joint and reported maximum ranges of motion of 9.3°and translations of 3.3mm for the fibula. The results show that the accuracy of the methodology is sufficient to quantify these small movements.

Conclusions

BVX combined with MRI can be used to accurately measure the syndesmosis joint. Future work will look at quantifying the accuracy of the talus to provide further understanding of normal ankle kinematics and to quantify the kinematics across a healthy population to act as a comparator for future patient studies.