Abstract

The reconstruction of peri-acetabular defects after severe bone loss or pelvic resection for tumor is among the most challenging surgical intervention. The Lumic® prosthesis (Implantcast, Buxtehude, Germany) was first introduced in 2008 in an effort to reduce the mechanical complications encountered with the classic peri-acetabular reconstruction techniques and to improve functional outcomes. Few have evaluated the results associated with the use of this recent implant.

A retrospective study from five Orthopedic Oncology Canadian centers was conducted. Every patient in whom a Lumic® endoprosthesis was used for reconstruction after peri-acetabular resection or severe bone loss with a minimal follow-up of three months was included. The charts were reviewed and data concerning patients’ demographics, peri-operative characteristics and post-operative complications was collected. Surgical and functional outcomes were also assessed.

Sixteen patients, 11 males and five females, were included and were followed for 28 months [3 – 60]. Mean age was 55 [17-86], and mean BMI reached 28 [19.6 – 44]. Twelve patients (75%) had a Lumic® after a resection of a primary sarcoma, two following pelvic metastasis, one for a benign tumor and one after a comminuted acetabular fracture with bone loss. Twelve patients (75%) had their surgery performed in one stage whereas four had a planned two-stage procedure. Mean surgical time was 555 minutes [173-1230] and blood loss averaged 2100 mL [500-5000]. MSTS score mean was 60.3 preoperatively [37.1 – 97] and 54.3 postoperatively [17.1-88.6]. Five patients (31.3%) had a cemented Lumic® stem. All patients got the dual mobility bearing, and 10 patients (62.5%) had the largest acetabular cup implanted (60 mm). In seven of these 10 patients the silver coated implant was used to minimize risk of infection. Five patients (31.3%) underwent capsular reconstruction using a synthetic fabric aiming to reduce the dislocation risk. Five patients had per-operative complications (31.3%), four were minor and one was serious (comminuted iliac bone fracture requiring internal fixation). Four patients dislocated within a month post-operatively and one additional patient sustained a dislocation one year post-operatively. Eight patients (50%) had a post-operative surgical site infection. All four patients who had a two-stage surgery had an infection. Ten patients (62.5%) needed a reoperation (two for fabric insertion, five for wash-outs, and three for implant exchange/removal). One patient (6.3%) had a septic loosening three years after surgery. At the time of data collection, 13 patients (81.3%) were alive with nine free of disease. Silver coating was not found to reduce infection risk (p=0.2) and capsuloplasty did not prevent dislocation (p=1).

These results are comparable to the sparse data published. Lumic® endoprosthesis is therefore shown to provide good functional outcomes and low rates of loosening on short to medium term follow-up. Infection and dislocation are common complications but we were unable to show benefits of capsuloplasty and of the use of silver coated implants. Larger series and longer follow-ups are needed.