Abstract

Over the last decade stemless shoulder arthroplasty has become increasingly popular. However, stability of metaphyseal loading humeral components remains a concern. This study aimed to assess the stability of the Affinis stemless humeral component using Radiostereometric analysis (RSA).

Patients underwent total shoulder arthroplasty via a standardised technique with a press-fit stemless humeral component and a cemented pegged glenoid. Tantalum beads were inserted into the humerus at the time of operation. RSA of the relaxed shoulder was completed at weeks 1, 6, 13, 26, 52 and 104 post-operatively. Stressed RSA with 12 newtons of abduction force was completed from week 13 onwards. ABRSA 5.0 software (Downing Imaging Limited, Aberdeen) was used to calculate humeral component migration and induced movement.

15 patients were recruited. Precision was: 0.041, 0.034, 0.086 and 0.101 mm for Superior, Medial, Posterior and Total Point Motion (TPM) respectively. The mean TPM over 2 years was 0.24 (0.30) mm, (Mean (Standard deviation)). The mean rate of migration per 3 month time period decreased from 0.45 (0.31) to 0.02 (0.01) mm over 2 years. Mean inducible movement TPM peaked at 26 weeks at 0.1 (0.08) mm, which reduced to 0.07 (0.06) mm by 104 weeks when only 3 patients had measurable inducible motion. There was no clear trend in direction of induced movement. There were no adverse events or revisions required.

We conclude migration of the humeral component was low with little inducible movement in the majority of patients implying initial and 2 year stability of the stemless humeral component.