THE EFFECTIVENESS OF FINE WIRE CLAMPING MECHANISM IN THE TAYLOR SPATIAL FRAME (TSF)

The British Limb Reconstruction Society (BLRS) 2021 Annual Scientific Meeting, Virtual Conference, held online, 15 April 2021.



Abstract

Introduction

In my paediatric Orthopaedic practice I use Kirchner wires for the fixation of the TSF on bone. I noted a significant percentage of wire loosening during the post-operative period. The aim of this project was to establish the effectiveness of the wire clamping mechanism and find ways to reduce the incidence of wire loosening when using the TSF.

Materials and Methods

In the first instance wire slippage was measured intra-operatively after the tensioner was removed using an intra-operative professional camera. Following this study mechanical tests were performed in the lab measuring the pull out properties of Kirchner wires using different bolts and different torque levels in order to tighten the wire on the fixator.

Results

Our clinical study confirmed wire slippage intra-operatively immediately after the tensioner was removed. Wire slippage after the tensioner was removed was found to vary from 0.01 mm to 0.51 mm (mean 0.19 mm). Our mechanical tests showed that the ideal torque for tightening the wire on the frame using a bolt was around 15 N.m. A comparison between cannulated and slotted bolts suggested that cannulated bolts are more effective as a clamping mechanism. A comparison between aluminium made Taylor Spatial frame rings and stainless steel made Ilizarov rings suggested that the Taylor Spatial frame rings are more effective as part of a clamping mechanism.

Conclusions

It is important that clinicians routinely measure the torque they use to clamp wires on circular external fixators. Clinicians and manufacturers are informed that the type of bolt used is important in maintaining wire tension. Manufacturers should design the ideal bolt which effectively grips the wire without the risk of fracture. The Ilizarov frame clamping mechanism can be effectively used with the Taylor Spatial frame.